Customization: | Available |
---|---|
Brand: | Testo |
Customized: | Non-Customized |
Still deciding? Get samples of $ !
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Feature
Robust and light instrument for daily use
30 seconds sensor zeroing
Integrated condensate trap
Integrated magnets for easy fixing to burner
testo 310 incl. battery and calibration protocol for the measurement of O2, CO, hPa and °C; IR printer (0554 3100); probe 180 mm with cone; case; mains unit incl. cable; silicon hose for pressure measurement; particle filter 5 off; 2 rolls spare thermal paper for printer.
Technical Data:
Temperature - TC Type K (NiCr-Ni) | |
---|---|
Measuring range |
-20 to +100 °C |
Accuracy |
±1 °C |
Resolution |
0.1 °C |
Reaction time |
< 50 s |
Temperature (ambient temperature)
Temperature - TC Type J (Fe-CuNi) | |
---|---|
Measuring range |
0 to +400 °C |
Accuracy |
±1 °C (0 to +100 °C) ±1.5 % of mv (> 100 °C) |
Resolution |
0.1 °C |
Reaction time |
< 50 s |
Temperature (flue gas)
Flue gas O2 | |
---|---|
Measuring range |
0 to 21 Vol.% |
Accuracy |
±0.2 Vol.% |
Resolution |
0.1 Vol.% |
Reaction time t90 |
30 s |
Flue gas Draught | |
---|---|
Measuring range |
-20 to +20 hPa |
Accuracy |
±0.03 hPa (-3.00 to +3.00 hPa) ±1.5 % of mv (Remaining Range) |
Resolution |
0.01 hPa |
Flue gas degree of effectivity, Eta (calculated) | |
---|---|
Measuring range |
0 to 120 % |
Resolution |
0.1 % |
Flue gas loss (calculated) | |
---|---|
Measuring range |
0 to 99.9 % |
Resolution |
0.1 % |
Flue gas CO2 calculation (calculated from O2) | |
---|---|
Measuring range |
0 to CO2 max (Display range) |
Accuracy |
±0.2 Vol.% |
Resolution |
0.1 Vol.% |
Reaction time t90 |
< 40 s |
Pressure measurement | |
---|---|
Measuring range |
-40 to +40 hPa |
Accuracy |
±0.5 hPa |
Resolution |
0.1 hPa |
Flue gas CO (without H2-compensation) | |
---|---|
Measuring range |
0 to 4000 ppm |
Accuracy |
±20 ppm (0 to 400 ppm) ±5 % of mv (401 to 2000 ppm) ±10 % of mv (2001 to 4000 ppm) |
Resolution |
1 ppm |
Reaction time t90 |
60 s |
Ambient CO | |
---|---|
Measuring range |
0 to 4000 ppm |
Accuracy |
±20 ppm (0 to 400 ppm) ±5 % of mv (401 to 2000 ppm) ±10 % of mv (2001 to 4000 ppm) |
Resolution |
1 ppm |
Reaction time |
60 s |
General technical data | |
---|---|
Dimensions |
201 x 83 x 44 mm |
Operating temperature |
-5 to +45 °C |
Display type |
LCD |
Display function |
Backlit 2-line display |
Power supply |
Battery: 1500 mAh, mains unit 5V/1A |
Storage temperature |
-20 to +50 °C |
Weight |
(with probe) Approx. 700 g |
1. Where is your company?
we are located in Guangzhou, the capital city of Guangdong Province, near to Shenzhen, Foshan & Hong Kong, very convenient for shipping.
2. How about the MOQ?
1 unit is ok.
3. Shipping methods and lead time?
By Express: 5-7 working days to your door (DHL, UPS, TNT, FedEx...).
By Air: 7-10 working days to your airport.
By Sea: Pls advise your port of destination, the exact days will be confirmed by our forwarders, and the following lead time is for your reference.Europe and America (25 - 35 days), Asia (3-7 days), Australia ( 16-23 days).
4. What is the term of Payment?
T/T, L/C,Western Union, Paypal;
30% T/T in advance,70% before delivery.
5. What are your main products?
Compressor;
Refrigeration gas;
Refrigeration spare parts ;
Cold room panels and equipments .
6. Packaging & Shipping:
Packing: wooden pallet, wooden case or with outer carton, or as per customers' specific requirements.